6525-00-853-8121
X-Ray Apparatus,Radiographic and
1 parts / 1 mfrs
| Part Number | CAGE | Manufacturer | RNCC | RNVC |
|---|---|---|---|---|
| FEDGGX00635 | 81348 | Federal Specifications Promulgated By General Services Administration Washington Dc | 4 | 1 |
Federal Specifications Promulgated By General Services Administration Washington Dc
CAGE 81348DEMIL
A
HMIC
P
PMIC
A
ESD/EMI
| MRC | Requirement | Reply |
|---|---|---|
| AAXX | MOUNTING TYPE | STATIONARY |
| AKCV | DRIVE TYPE | MOTOR |
| AKYD | ACCESSORY COMPONENTS AND QUANTITY | PHOTO-TIMING CIRCUIT 1,COMPRESSION DEVICE 1,MULTIDIAPHRAGM ADJUSTABLE RADIOGRAPHIC CONE 1,CONTROL /SEPARATE KILOVOLTAGE/FOR FLUOROSCOPIC KILOVO LTAGES 1,FOOTREST 1,MILLIAMPERE-SECOND METER 1,SPOT-FILM CHANGE-OVER DEVICE 1,MYLOGRAPHIC SUPPORT PATIENT 1,SHOULDER SUPPORT 1 |
| APGF | DESIGN TYPE | STATIONARY |
| APHE | OPERATION METHOD | AUTOMATIC |
| AQZF | CONTROL TYPE | VERTICAL |
| ARNF | TUBE QUANTITY | 2 |
| ARQE | ELECTRIC POWER RATING | 65000.0 WATTS NOMINAL |
| AWGH | TUBE TYPE | DOUBLE FOCUS, ROTATING ANODE RADIOGRAPHIC AND DOUBLE FOCUS, ROTATING ANODE FLUOROSCOPIC |
| BNBB | TABLE TYPE | TILTING |
| BPKZ | RECTIFICATION TYPE | FULL-WAVE |
| CCFN | SCREEN TYPE | W/SPOT-FILM |
| CLPG | TUBE COOLING METHOD | AIR |
| CLPH | TUBE MOUNTING TYPE | TUBE STAND AND TABLE |
| CLPR | TUBE STAND MOUNTING TYPE | OVERHEAD |
| CLPT | TILT RANGE | 90 DEG TRENDELENBURG TO VERTICAL |
| CLPW | BUCKY DIAPHRAGM | INCLUDED |
| CLPX | RADIOGRAPHIC TIMER TYPE | EXPOSURE |
| CLPY | FLUOROSCOPIC TIMER TYPE | SYNCHRONOUS MOTOR |
| CLQD | TRANSFORMER RECTIFICATION METHOD | ELECTRON TUBE |
| CLQF | TRANSFORMER OUTPUT CAPACITY | 500 MA AT 125 KV PEAK,300 MA AT 150 KV PEAK |
| CMGN | SCREEN LENGTH | 12.000 INCHES NOMINAL |
| CMGP | SCREEN WIDTH | 12.000 INCHES NOMINAL |
| CMGQ | CASSETTE SIZE | 8 IN. BY 10 IN. |
| CTJL | FLUOROSCOPIC SCREEN TYPE | ANY ACCEPTABLE |
| CXCY | PART NAME ASSIGNED BY CONTROLLING AGENCY | X-RAY APPARATUS RADIOGR-FLUOROSCOPIC 500MA 150PKV 230V 60HZ AC |
| ELEC | VOLTAGE IN VOLTS | 190.0 AND 250.0 |
| FAAZ | PHASE | SINGLE |
| FEAT | SPECIAL FEATURES | REQUISITION SCHOULD STATE WHETHER HIGH-SPEED INCERTS ARE REQUIRED |
| FREQ | FREQUENCY IN HERTZ | 60.0 |
| SUPP | SUPPLEMENTARY FEATURES | REQUISITION SHOULD STATE FLOOR-LOADING CAPACITY,CEILING STRENGTH AND HEIGHT,WALL AND FLOOR CONSTRUCTION (CONCRETE,WOOD,ETC),ELECTRICAL SUPPLY AND CAPACITY,DIAGRAM OF ROOM INDICATING DIMENSIONS,COMPATABILITY W/EXISTING FACILITIES,WHETHER OR NOT CONTRACTOR INSTALLATION SERVICE IS REQUIRED,AND WHETHER HIGH-SPEED INCERTS ARE REQUIRED |
Need full procurement data, packaging, and standardization info?
View Complete NSN Data on NSN LookupQuick Reference
About NSN 6525-00-853-8121
NSN 6525-00-853-8121 is a National Stock Number assigned to X-Ray Apparatus,Radiographic and. This NSN part number belongs to Federal Supply Class 6525 (IMAGING EQUIPMENT AND SUPPLIES: MEDICAL, DENTAL, VETERINARY) under Federal Supply Group 65 (MEDICAL, DENTAL, AND VETERINARY EQUIPMENT AND SUPPLIES). There are 1 part number associated with this NSN from 1 manufacturer.
To find the part number for NSN 6525-00-853-8121, refer to the part number cross-reference table above. Each part number is linked to its manufacturer through a CAGE (Commercial and Government Entity) code.